Basic Pacing ConceptsPart III
Slide 2 :
Sensing
Slide 3 :
Sensing Sensing is the ability of the pacemaker to “see” when a natural (intrinsic) depolarization is occurring Pacemakers sense cardiac depolarization by measuring changes in electrical potential of myocardial cells between the anode and cathode
Slide 4 :
An Electrogram (EGM) is the Recording of Cardiac Waveforms Taken From Within the Heart Intrinsic deflection on an EGM occurs when a depolarization wave passes directly under the electrodes Two characteristics of the EGM are: Signal amplitude Slew rate
Slide 5 :
Slew Rate of the EGM Signal Measures the Change in Voltage with Respect to the Change in Time The longer the signal takes to move from peak to peak: The lower the slew rate The flatter the signal Higher slew rates (number in mV) translate to greater sensing Measured in volts per second Voltage Time Slope Slew rate= Change in voltage Time duration of voltage change
Slide 6 :
A Pacemaker Must Be Able to Sense and Respond to Cardiac Rhythms Accurate sensing enables the pacemaker to determine whether or not the heart has created a beat on its own The pacemaker is usually programmed to respond with a pacing impulse only when the heart fails to produce an intrinsic beat
Slide 7 :
Accurate Sensing... Ensures that undersensing will not occur –the pacemaker will not miss P or R waves that should have been sensed Ensures that oversensing will not occur – the pacemaker will not mistake extra-cardiac activity for intrinsic cardiac events Provides for proper timing of the pacing pulse – an appropriately sensed event resets the timing sequence of the pacemaker
Slide 8 :
Undersensing . . . Pacemaker does not “see” the intrinsic beat, and therefore does not respond appropriately Intrinsic beat not sensed Scheduled pace delivered VVI / 60
Slide 9 :
Oversensing An electrical signal other than the intended P or R wave is detected Marker channel shows intrinsic activity... ...though no activity is present
Slide 10 :
Sensitivity – The Greater the Number, the Less Sensitive the Device to Intracardiac Events
Slide 11 :
Sensitivity Amplitude (mV) Time 5.0 2.5 1.25
Slide 12 :
Sensitivity Amplitude (mV) Time 5.0 2.5 1.25
Slide 13 :
Sensitivity Amplitude (mV) Time 5.0 2.5 1.25
Slide 14 :
Accurate Sensing Requires That Extraneous Signals Be Filtered Out Sensing amplifiers use filters that allow appropriate sensing of P waves and R waves and reject inappropriate signals Unwanted signals most commonly sensed are: T waves Far-field events (R waves sensed by the atrial channel) Skeletal myopotentials (e.g., pectoral muscle myopotentials)
Slide 15 :
Accurate Sensing is Dependent on . . . The electrophysiological properties of the myocardium The characteristics of the electrode and its placement within the heart The sensing amplifiers of the pacemaker
Slide 16 :
Factors That May Affect Sensing Are: Lead polarity (unipolar vs. bipolar) Lead integrity Insulation break Wire fracture EMI – Electromagnetic Interference
Slide 17 :
Unipolar Sensing Produces a large potential difference due to: A cathode and anode that are farther apart than in a bipolar system _
Slide 18 :
Bipolar Sensing Produces a smaller potential difference due to the short interelectrode distance Electrical signals from outside the heart such as myopotentials are less likely to be sensed
Slide 19 :
An Insulation Break May Cause Both Undersensing or Oversensing Undersensing occurs when inner and outer conductor coils are in continuos contact Signals from intrinsic beats are reduced at the sense amplifier and amplitude no longer meets the programmed sensing value Oversensing occurs when inner and outer conductor coils make intermittent contact Signals are incorrectly interpreted as P or R waves
Slide 20 :
A Wire Fracture Can Cause Both Undersensing and Oversensing Undersensing occurs when the cardiac signal is unable to get back to the pacemaker – intrinsic signals cannot cross the wire fracture Oversensing occurs when the severed ends of the wire intermittently make contact, which creates potentials interpreted by the pacemaker as P or R waves
Slide 21 :
Electromagnetic Interference
Slide 22 :
Electromagnetic Interference (EMI) Interference is caused by electromagnetic energy with a source that is outside the body Electromagnetic fields that may affect pacemakers are radio-frequency waves 50-60 Hz are most frequently associated with pacemaker interference Few sources of EMI are found in the home or office but several exist in hospitals
Slide 23 :
EMI May Result in the Following Problems: Oversensing Transient mode change (noise reversion) Reprogramming (Power on Reset or “POR”)
Slide 24 :
Oversensing May Occur When EMI Signals Are Incorrectly Interpreted as P Waves or R Waves Pacing rates will vary as a result of EMI: Rates will accelerate if sensed as P waves in dual-chamber systems (P waves are “tracked”) Rates will be low or inhibited if sensed in single-chamber systems, or on ventricular lead in dual-chamber systems
Slide 25 :
EMI “Noise” sensed by the pacemaker Should have paced
Slide 26 :
Noise Reversion VP VP SR SR SR SR Noise Sensed VVI/60 Continuous refractory sensing will cause pacing at the lower or sensor driven rate
Slide 27 :
EMI May Lead to Inadvertent Reprogramming of the Pacing Parameters Device will revert to Power on Reset (POR or “backup” mode) Power on Reset may exhibit a mode and rate change which are often the same as ERI In some cases, reprogrammed parameters may be permanent
Slide 28 :
New technologies will continue to create new, unanticipated sources of EMI: Cellular phones (digital)
Slide 29 :
Sources of EMI Are Found Most Commonly in Hospital Environments Sources of EMI that interfere with pacemaker operation include surgical/therapeutic equipment such as: Electrocautery Transthoracic defibrillation Extracorporeal shock-wave lithotripsy Therapeutic radiation RF ablation TENS units MRI
Slide 30 :
Sources of EMI Are Found More Rarely in: Home, office, and shopping environments Industrial environments with very high electrical outputs Transportation systems with high electrical energy exposure or with high-powered radar and radio transmission Engines or subway braking systems Airport radar Airplane engines TV and radio transmission sites
Slide 31 :
Electrocautery is the Most Common Hospital Source of Pacemaker EMI Outcomes Oversensing–inhibition Undersensing (noise reversion) Power on Reset Permanent loss of pacemaker output (if battery voltage is low) Precautions Reprogram mode to VOO/DOO, or place a magnet over device Strategically place the grounding plate Limit electrocautery bursts to 1-second burst every 10 seconds Use bipolar electrocautery forceps
Slide 32 :
Transthoracic Defibrillation Outcome Inappropriate reprogramming of the pulse generator (POR) Damage to pacemaker circuitry Precautions Position defibrillation paddles apex-posterior (AP) and as far from the pacemaker and leads as possible
Slide 33 :
Magnetic Resonance Imaging (MRI) is Generally Contraindicated in Patients with Pacemakers Outcomes Extremely high pacing rate Reversion to asynchronous pacing Precautions Program pacemaker output low enough to create persistentnon-capture, ODO or OVO mode
Slide 34 :
Lithotripsy Shock Waves May Have an Effect on Pacemaker Systems Outcomes indual-chamber modes: Inhibition of ventricular pacing Outcomes in rate adaptive pacemakers High pacing rates Piezoelectric crystal damage Precautions: Program pacemaker to VVI or VOO mode Lithotriptor focal point should be greater than 6 inches from the pacemaker Carefully monitor heart function throughout procedure
Slide 35 :
Radiation Energy May Cause Permanent Damage Certain kinds of radiation energy may cause damage to the semi-conductor circuitry Ionizing radiation used for breast or lung cancer therapy Damage can be permanent and requires replacement of the pacemaker
Slide 36 :
Therapeutic Radiation May Cause Severe Damage Outcomes: Pacemaker circuit damage Loss of output “Runaway” Precautions: Keep cumulative radiation absorbed by the pacemaker to less than 500 rads; shielding may be required Check pacemaker after radiation sessions for changes in pacemaker function (can be done transtelephonically)
Slide 37 :
Pacemaker Features That Address Interference Pacemaker sensing circuits amplify, filter and either process or reject incoming signals
Slide 38 :
Rate Responsive Pacing
Slide 39 :
Rate Response Rate responsive (also called rate modulated) pacemakers provide patients with the ability to vary heart rate when the sinus node cannot provide the appropriate rate Rate responsive pacing is indicated for: Patients who are chronotropically incompetent (heart rate cannot reach appropriate levels during exercise or to meet other metabolic demands) Patients in chronic atrial fibrillation with slow ventricular response
Slide 40 :
Rate Responsive Pacing Cardiac output (CO) is determined by the combination of stroke volume (SV) and heart rate (HR) SV X HR = CO Changes in cardiac output depend on the ability of the HR and SV to respond to metabolic requirements
Slide 41 :
Rate Responsive Pacing SV reserves can account for increases in cardiac output of up to 50% HR reserves can nearly triple total cardiac output in response to metabolic demands
Slide 42 :
Rate Responsive Pacing When the need for oxygenated blood increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output Adjusting Heart Rate to Activity Normal Heart Rate Rate Responsive Pacing Fixed-Rate Pacing Daily Activities
Slide 43 :
A Variety of Rate Response Sensors Exist Those most accepted in the market place are: Activity sensors that detect physical movement and increase the rate according to the level of activity Minute ventilation sensors that measure the change in respiration rate and tidal volume via transthoracic impedance readings
Slide 44 :
Rate Responsive Pacing Activity sensors employ a piezoelectric crystal that detects mechanical signals produced by movement The crystal translates the mechanical signals into electrical signals that in turn increase the rate of the pacemaker Piezoelectric crystal
Slide 45 :
Rate Responsive Pacing Minute Ventilation (MV) is the volume of air introduced into the lungs per unit of time MV has two components: Tidal volume–the volume of air introduced into the lungs in a single respiration cycle Respiration rate–the number of respiration cycles per minute
Slide 46 :
Rate Responsive Pacing Minute ventilation can be measured by measuring the changes in electrical impedance across the chest cavity to calculate changes in lung volume over time
Slide 47 :
Pacing systems Electrical concepts Stimulation thresholds Sensing Electromagnetic Interference (EMI) Rate response Summary of Basic Pacing Concepts Module
Slide 48 :
General Medtronic Pacemaker Disclaimer INDICATIONS Medtronic pacemakers are indicated for rate adaptive pacing in patients who may benefit from increased pacing rates concurrent with increases in activity (Thera, Thera-i, Prodigy, Preva and Medtronic.Kappa 700 Series) or increases in activity and/or minute ventilation (Medtronic.Kappa 400 Series). Medtronic pacemakers are also indicated for dual chamber and atrial tracking modes in patients who may benefit from maintenance of AV synchrony. Dual chamber modes are specifically indicated for treatment of conduction disorders that require restoration of both rate and AV synchrony, which include various degrees of AV block to maintain the atrial contribution to cardiac output and VVI intolerance (e.g., pacemaker syndrome) in the presence of persistent sinus rhythm. 9790 Programmer The Medtronic 9790 Programmers are portable, microprocessor based instruments used to program Medtronic implantable devices. 9462 The Model 9462 Remote Assistant™ is intended for use in combination with a Medtronic implantable pacemaker with Remote Assistant diagnostic capabilities. CONTRAINDICATIONS Medtronic pacemakers are contraindicated for the following applications: · Dual chamber atrial pacing in patients with chronic refractory atrial tachyarrhythmias. · Asynchronous pacing in the presence (or likelihood) of competitive paced and intrinsic rhythms. · Unipolar pacing for patients with an implanted cardioverter-defibrillator because it may cause unwanted delivery or inhibition of ICD therapy. · Medtronic.Kappa 400 Series pacemakers are contraindicated for use with epicardial leads and with abdominal implantation. WARNINGS/PRECAUTIONS Pacemaker patients should avoid sources of magnetic resonance imaging, diathermy, high sources of radiation, electrosurgical cautery, external defibrillation, lithotripsy, and radiofrequency ablation to avoid electrical reset of the device, inappropriate sensing and/or therapy. 9462 Operation of the Model 9462 Remote Assistant™ Cardiac Monitor near sources of electromagnetic interference, such as cellular phones, computer monitors, etc. may adversely affect the performance of this device. See the appropriate technical manual for detailed information regarding indications, contraindications, warnings, and precautions. Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Slide 49 :
Medtronic Leads For Indications, Contraindications, Warnings, and Precautions for Medtronic Leads, please refer to the appropriate Leads Technical Manual or call your local Medtronic Representative. Caution: Federal law restricts this device to sale by or on the order of a Physician. Note: This presentation is provided for general educational purposes only and should not be considered the exclusive source for this type of information. At all times, it is the professional responsibility of the practitioner to exercise independent clinical judgment in a particular situation.
There is no comments now for this Slide.
Post a comment
Post Comment on Twitter
Post Comment on SlideWorld
Comments:
Subscribe to follow-up comments
SlideWorld will not store your password. SlideWorld will maintain your privacy.
Twitter Username:
Twitter Password:
Comments:
Email:
Subscribe to follow-up comments
review of basic echo...
Interventional Pharm...
Dental Caries curren...
Community Acquired P...
Pulmonary Function T...
ATP III Guidelines
Free Powerpoint Templates
SlidesOnline
4 Years ago.
Category:
General
Tags:
other
3795 Views, 0 favourite
Medtronic pacemakers are indicated for rate adaptive pacing in patients who may .... In VVI pacing,